Comparison of Venous and Capillary Whole Blood Estimations of Covid-19 IgG and IgM Antibodies

ABSTRACT

The testing of Covid-19 antibodies is fast becoming an important activity in the understanding and management of Covid-19 infection. Most of the tests are conducted on the venous blood samples. As quite a number of the tests being produced are point of care tests such as the lateral flow methods, finger prick or capillary samples are more convenient than venous samples. A prospective observational study was carried out on paired capillary and venous samples from 41 healthy blood donors using the Boditech iCHROMA FIA fluorescence immunoassay (FIA) method for measuring IgM and IgG. Whilst there was agreement between the IgM results of 31/41 (76%) of the capillary and venous blood samples, with a disagreement between the results of 10/41 (24%) of the capillary and venous blood samples. There was agreement between the IgG results of 39/41 (95%) of the capillary and venous blood samples, with a disagreement between the results of 2/41 (5%) of the capillary and venous blood samples. In conclusion, using capillary samples as an alternative method to venous whole blood for estimating and measuring the Covid-19 IgG antibodies are acceptable, whilst IgM antibody testing using capillary samples warrants careful interpretation.

INTRODUCTION

The detection of antibodies has been important in understanding the course of the Covid-19 infection. There have been a number of methods such as enzyme linked immunoassay (ELISA), chemiluminescence immunoassays (CMIA) using discrete and bench top analysers, we haves recently compared the Boditech iCHROMA a portable fluorescent immunoassay (FIA) method of measurement of Covid-19 IgG antibody with the Abbott Architect SARS-CoV-2 and found very good overall agreement of 95%, with a sensitivity of 100% and a specificity of 90% [1]. In addition, the Boditech iCHROMA was shown to have performed very well with external quality control material provided by the UK National Institute of Biological Standards and Control (NIBSC) [2]. Most of the tests have been conducted on serum or plasma from venous whole blood samples. Several the tests being produced are point of care tests such as the lateral flow methods require finger prick or capillary samples which are more convenient than venous samples this sampling. In a study, where a Covid-19 IgG enzyme linked immunoassay (ELISA) antibody assay was studied in plasma from venous blood samples that had been collected in lithium heparin tubes and matched with capillary blood samples collected in lithium heparin tubes, there was a very high measure of concordance (Cohen’s kappa coefficient of >0.88) between the two sample groups, indicating that capillary blood sampling is a reliable venous blood sample alternative [3]. In another study, fingerstick plasma showed a 100% concordance (R2 = 0.997) with matched patient venous plasma using the Abbott Architect™ SARSCoV- 2 IgG assay (Anderson et al). In this prospective observational study on paired capillary and venous whole blood samples from 41 healthy blood donors, we studied using the Boditech iCHROMA fluorescence immunoassay (FIA) method for measuring Covid-19 IgM and IgG antibodies to determine the practical solution of using capillary whole blood samples to venous whole blood.

MATERIALS AND METHODS

Study Participants

Forty-one subjects (23 males and 18 females aged between 19 and 57 years) from a blood donor panel obtained from Cambridge Biosciences were recruited to volunteer for this study. Cambridge Biosciences obtains fresh human blood service in partnership with London-based Research Donors, which is a HTA licensed and ISO 9001 2015 certified company with Research Ethics (REC) approval as a Research Tissue bank. All donors provided consented to use their blood samples for research. Donors of Human Biospecimens supplied by Cambridge Bioscience have consented to the use of their samples for biological research purposes.

Venous and Capillary Blood Collection

Trained staff at London based research donors collected approximately 10ml of venous whole blood into blood collection tubes (BD Vacutainer® treated with lithium heparin). Samples were aliquoted from the tube for IgG and IgM analysis using the Boditech iCHROMA method described later. Capillary blood samples were also collected from the volunteers by the staff into lithium heparin coated Microvette® 100 by Sarstedt. Samples were aliquoted from the tube for IgG and IgM analysis using the Boditech iCHROMA method described later.

Boditech iCHROMA Method Principle: The test uses a sandwich immunodetection method; fluorescence labelled conjugates in a dried detection buffer binds to antibody in sample, forming antibody-antigen complexes, and migrates onto nitrocellulose matrix to be captured by the other immobilized anti-human IgG on test strip. The more antigen-antibody complexes lead to stronger fluorescence signal by the detector antigen which is processed by iCHROMA. The iCHROMA processes the signal using a cut off index of 0.9-1.1, results <0.9 are interpreted as negative, results between 0.9 and 1.1 are interpreted as indeterminate and results >1.1 are interpreted as positive.

The paired capillary and venous samples were analysed using the Boditech iCHROMA Fluorescence Immunoassay (FIA) IgG assay described below:

a) Transfer 150μL of detector diluent using a pipette into the detector tube containing a granule. When the granule is completely dissolved it becomes the detection buffer.
b) Aspirate 10μL of whole blood/serum/plasma/control with a pipette, and add into the detector tube, close and shake the tube at least 10 times.
c) Pipette out 75μL of the content of the tube and load it into the sample well on the test cartridge and leave or 10 minutes.
d) Insert the test cartridge into the cartridge holder in iCHROMA II device and press start
e) Read the result on the display screen of the iCHROMA II device.

RESULTS

A total of 41 volunteers took part in this study, of which 23 (56%) were male and 18 (44%) females. The average age was 34 years with a range of 19-57 years.

Covid-19 IgG

Covid-19 IgG antibodies were estimated on 41 subjects venous whole blood samples collected in lithium heparin tubes, 14 of the 41 (34%) subject’s samples were identified as Covid-19 IgG positive, 2 of the 41 (5%) as indeterminate and 25 of the 41 (61%) as Covid-19 IgG negative. Covid-19 IgG antibodies were estimated in 41 subject’s capillary whole samples collected in lithium heparin tubes, 14 of the 41 (31%) subject’s samples were identified as Covid-19 IgM positive, none of the 41 (2.4%) as indeterminate and 27 of the 41 (966%) as Covid-19 IgM negative (Table 1).

The minimum, mean and maximum cut off indices of Covid-19 IgG detected in the venous whole blood samples were 0.00, 5.434 and 23.30 respectively, compared to minimum, mean and maximum cut off indices of Covid-19 IgG detected in the capillary whole blood samples was 0.00, 6,64 and 41.60 respectively. Paired T tests showed there was no significant difference (p=0.1968) between means of the IgG antibodies detected in the venous and the capillary whole blood samples (Figure 1).

There was concordance between the results of 39/41 (95%) of the capillary and venous blood samples, with a discordance between the results of 2/41 (5%) of the capillary and venous blood samples. The discordant samples (3643 and 3592), the cut off indices in the venous blood samples were 1.00, with the interpretation of indeterminate indicating a very low level of IgG antibodies were detected, their corresponding capillary blood sample was neagative (Table 2).

Covid-19 IgM

Covid-19 IgM antibodies were estimated on 41 subjects venous whole blood samples collected in lithium heparin tubes, 11 of the 41 (27%) subject’s samples were identified as Covid-19 IgM positive, 3 of the 41 (7%) as indeterminate and 27 of the 41 (66%) as Covid-19 IgM negative. Covid-19 IgM antibodies were estimated in 41 subject’s capillary whole samples collected in lithium heparin tubes, 3 of the 41 (7.3%) subject’s samples were identified as Covid-19 IgM positive, 1 of the 41 (2.4%) as indeterminate and 37 of the 41 (90.3%) as Covid-19 IgM negative.

The minimum, mean and maximum cut off indices of Covid-19 IgM detected in the venous whole blood samples were 0.00, 0.9588 and 5.20 respectively, compared to minimum, mean and maximum cut off indices of Covid-19 IgM detected in the capillary whole blood samples was 0.00, 0.4195 and 4.00 respectively. Paired T tests showed a significant difference (p<0.0001) between means of the IgM antibodies detected in the venous and the capillary whole blood samples (Figure 2).

There was concordance between the results of 31 of the 41 (76%) of the capillary blood and venous blood samples, with a discordance between the results of 10/41 (24%) of the capillary and venous blood samples. The discordant venous blood samples were indeterminate or positive with very low cut off indices (Table 3,4).

DISCUSSION

In this study, there was a high level of agreement (95%) between the detection of Covid-19 IgG antibodies in venous and capillary whole blood samples. The discordance (5%) was seen in two venous whole blood samples that were interpreted as indeterminate indicating the presence of a low level of Covid-19 IgG antibodies in the venous whole blood samples the corresponding capillary blood samples were undetected and interpreted as negative and from the cut off indices had very low levels of Covid-19 IgG antibodies. The Covid-19 IgG antibody results in this study are consistent with findings in other studies, in which there was a concordance of >94% and a discordance of 2.4% between capillary and venous blood samples for Covid-19 IgG using an ELISA method [3]. Anderson et al. [4] also showed 100% concordance between fingerstick plasma and venous plasma on the Architect SARS-CoV-2 IgG assay.

There was a lower level of agreement (75%) for Covid-19 IgM antibodies between venous and capillary blood samples. This is much lower than what has been described by Anderson et al, where the concordance between plasma and venous or fingerstick dried bloods spots were 98.4% and 100%, respectively (Anderson et al). There is not much in the literature looking at the concordance of Covid-19 IgM antibodies between capillary and venous samples. A possible explanation for the higher discordance of IgM antibodies observed between the venous and capillary whole blood sample compared to IgG antibodies, could be as IgM is a larger molecule (molecular weight of 900kDa), compared to IgG which is much smaller molecule (molecular weight of 150kDa) and that is consistent with the results seen in this study, in which Covid-19 IgM antibodies were detected in 10 venous blood samples but not in the corresponding capillary blood samples [5-8].

CONCLUSION

Our results indicate that capillary blood self-sampling is a reliable and feasible alternative to venepuncture for serological assessment in Covid-19 IgG antibodies and that the assessment of Covid-19 IgM antibodies be taken with some degree of caution.

REFERENCES

  1. Bass M, Bolodeoku J, Stevenson E, Anyaeche C, Kim TK, et al. (2020) Agreement of the point of care test (POCT) boditech iCHROMA Covid-19 IgG antibody assay with the abbott architect SARS-CoV-2 IgG antibody assay. Ann Immunol Immunther 2 (2): 000121.
  2. Bolodeoku J, Bass M, Anyaeche C, Kim TK, Retnasingham V (2020) Performance of the Boditech iCHROMA Covid 19 IgG antibody assay with the external quality control from UK NIBSC (National Institute of Biological Standards and Control). J Clin Med Rev and Rep 4(1): 1-4.
  3. Brown L, Byrne RL, Fraser A, Owen SI, Cubas-Atienzar AI, et al. (2021) Self-sampling of capillary blood for SARS-CoV-2 serology. Sci Rep 11: 7754.
  4. Anderson M, Holzmayer V, Vallari A, Taylor R, Moy J, et al. (2021) Expanding access to SARS-CoV-2 IgG and IgM serologic testing using fingerstick whole blood, plasma, and rapid lateral flow assays. Journal of clinical Virology 141: 104855.
  5. Bolodeoku J, Bass M, Anyaeche C, Retnasingham V (2021) A mild case of COVID-19 infection: An observational longitudinal study 27 days post symptom of antigen, antibodies (IgM & IgG), IL-6 and D-Dimer. A review. Am J Biomed Sci & Res 11(6): 519-523.
  6. Muhammad M, Inna S, Meghan M, Paul N, Navdeep C, et al. (2021) Comparative analysis of capillary vs venous blood for serologic detection of SARS-CoV-2 antibodies by RPOC lateral flow tests. Open Forum Infectious Diseases 8(3): 1-4.
  7. Krleza JL, Dorotic A, Grzunov A, Maradin M (2015) Croatian society of medical biochemistry and laboratory medicine. Capillary blood sampling: national recommendations on behalf of the Croatian Society of Medical Biochemistry and Laboratory Medicine. Biochem Med 25(3): 335-358.
  8. Kupke IR, Kather B, Zeugner S (1981) On the composition of capillary and venous blood serum. Clin Chim Acta. 112(2): 177-185.

Article Type

Research Article

Publication history

Received Date: May 24, 2022
Published: June 20, 2022

Address for correspondence

J Bolodeoku, JB Consulting MDP Ltd, UK

Copyright

©2022 Open Access Journal of Biomedical Science, All rights reserved. No part of this content may be reproduced or transmitted in any form or by any means as per the standard guidelines of fair use. Open Access Journal of Biomedical Science is licensed under a Creative Commons Attribution 4.0 International License

How to cite this article

J Bolodeoku, M Bass, C Anyaeche, V Retnasinhgham. Comparison of Venous and Capillary Whole Blood Estimations of Covid-19 IgG and IgM Antibodies. 2022- 4(3) OAJBS.ID.000460.

Author Info

J Bolodeoku1*, M Bass1, C Anyaeche2 and V Retnasinhgham1

1Path Direct Laboratory, UK
2Pathaway Services Limited, UK

Figure 1: Showing cut off indices of Covid-19 IgG samples (capillary and venous whole blood).

oajbs-G460-1

Figure 2: Showing cut off indices of Covid-19 IgM samples (capillary and venous whole blood).

oajbs-G460-2

Table 1: Showing agreement of venous and capillary whole blood samples with Boditech iCHROMA Covid-19 IgG interpretation (positive, indeterminate, negative).

oajbs-T460-1

Table 2: Showing concordance of result interpretation of Boditech iCHROMA Covid-19 IgG using venous and capillary whole blood samples.

oajbs-T460-2

Table 3: Showing agreement of venous and capillary whole blood samples with Boditech iCHROMA Covid-19 IgM interpretation (positive, indeterminate, negative).

oajbs-T460-3

Table 4: Showing concordance of result cut off indices and interpretation of Boditech iCHROMA Covid-19 IgM using venous and capillary whole blood samples.

oajbs-T460-4